Item Occupational exposure to staff during endoscopic retrograde cholangiopancreatography in Sudan(Oxford University Press, 2011) Sulieman, A; Elzaki, M; Khalil, MohamedEndoscopic retrograde cholangiopancreatography (ERCP) procedure is an invasive technique that requires fluoroscopic and radiographic exposure. The purpose of this study was to determine the occupational dose of ionising radiation at three gastroenterology departments (Fedial, Soba and Ibn seena hospitals) in Khartoum, Sudan. The radiation dose was measured during 55 therapeutic ERCP procedures. Thermoluminescence dosemeters were used. The mean radiation dose for the first operator was 0.27 mGy for the eye lens, 0.21 for the thyroid, 0.32 for the chest, 0.17 for the hand and 0.22 for the leg. The mean radiation dose for the second operator was 0.21 mGy for the hand and 0.20 mGy for the chest, while the mean radiation dose for the nurse was 0.44 mGy for the hand and 0.19 for the chest. The radiation dose received by the staff in these hospitals was found to be higher than most of the values in the literature. The radiation absorbed dose received by the different organs is relatively low. Additional studies need to be conducted for radiation dose optimisation.