Saeed, HaithamAli, Ahmed M. A.Elberry, Ahmed A.Eldin, Abeer SalahRabea, HodaAbdelrahim, Mohamed E. A.2019-11-092019-11-092018-06Cited References in Web of Science Core Collection: 381094-5539https://doi.org/10.1016/j.pupt.2018.04.005https://www.ncbi.nlm.nih.gov/pubmed/29635073Accession Number: WOS:000435057500008Backgrounds: Substituting nebulisers by another, especially in non-invasive ventilation (NIV), involves many process-variables, e.g. nebulizer-type and fill-volume of respirable-dose, which might affect patient optimum therapy. The aim of the present work was to use neural-networks and genetic-algorithms to develop performance-models for two different nebulizers. Methods: In-vitro, ex-vivo and in-vivo models were developed using input-variables including nebulizer-type [jet nebulizer (JN) and vibrating mesh nebulizer (VMN)] fill-volumes of respirable dose placed in the nebulization chamber with an output-variable e.g. average amount reaching NIV patient. Produced models were tested and validated to ensure effective predictivity and validity in further optimization of nebulization process. Results: Data-mining produced models showed excellent training, testing and validation correlation-coefficients. VMN showed high nebulization efficacy than JN. JN was affected more by increasing the fill-volume. The optimization process and contour-lines obtained for in-vivo model showed increase in pulmonary-bioavailability and systemic-absorption with VMN and 2 mL fill-volumes. Conclusions: Modeling of aerosol-delivery by JN and VMN using different fill-volumes in NIV circuit was successful in demonstrating the effect of different variable on dose-delivery to NIV patient. Artificial neural networks model showed that VMN increased pulmonary-bioavailability and systemic-absorption compared to JN. VMN was less affected by fill-volume change compared to JN which should be diluted to increase delivery.enNon-invasive ventilationModeling; NebulizerNeural networksFill volumeARTIFICIAL NEURAL-NETWORKSLUNG FOLLOWING INHALATIONNEXT-GENERATION IMPACTORMETERED-DOSE INHALERDRY POWDER INHALERIN-VITROMECHANICAL VENTILATIONAERODYNAMIC CHARACTERISTICSSYSTEMIC BIOAVAILABILITYAEROSOL DELIVERYModeling and optimization of nebulizers' performance in non-invasive ventilation using different fill volumes: Comparative study between vibrating mesh and jet nebulizersArticlehttps://doi.org/10.1016/j.pupt.2018.04.005